Casella-cel CEL-368 Manual de usuario Pagina 8

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 99
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 7
displays the current status of the data stores on a MEMORY USE
screen. This allows the user to see if there is space for the next run, or
whether the contents of a store must first be cleared to make room as
described in Section 1.6. Then the user is offered the choice of recali-
brating (i.e performing a field accuracy check) or continuing with the ex-
isting calibration settings. If the battery level is too low for reliable
operation, a warning is displayed.
1.4 Field Accuracy Check (Acoustic Calibration)
Either: Press to use the existing calibration settings and proceed
directly to the SETUP menu (Section 1.5).
Or: Press to perform a field accuracy check (acoustic calibration).
Acoustic calibration of the CEL-268 will normally be performed with a
CEL-284/2 Acoustic Calibrator (IEC 942 Class 1L), while the CEL-368
will use a CEL-282 Acoustic Calibrator IEC 942 Class 2L).
With the CEL-268 and CEL-284/2, carefully fit the calibrator over the
end of the microphone, making sure the microphone is correctly seated
against the shoulder in the calibrator cavity.
EXIT
STOP
CAL
Page 4 - CEL-268/368 Operator's Handbook
Vista de pagina 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 98 99

Comentarios a estos manuales

Sin comentarios