CEL-63x Environmental Noise Meter
71 Casella CEL
9 Additional information
The following additional information relates to testing under IEC 61672-1:2002
section 9.3
1. Reference sound pressure level
The reference sound pressure level can be selected to 94 dB or 114 dB.
2. Reference level range
The CEL-63x instrument has a single measurement range of 0 dB to 140 dB.
3. Microphone reference point
The microphone reference point is the centre of the microphone‟s diaphragm.
The 0° reference direction is perpendicular to the microphone‟s diaphragm.
4. Acoustical frequency response testing
Table 3 shows the pressure to free-field correction data for the CEL-251
microphone used for periodic testing.
Table 3. Pressure to 0° free-field corrections
Expanded
uncertainty of
corrections at 95 %
probability (k=2)
Bruel & Kjear 4226
calibrator
Bruel & Kjear UA0033
electrostatic actuator