
Options for the COM port, baud rate and Stop bits can now be
configured.
4. Select the COM port to which the CEL-450/490 is connected.
5. Select the same baud rate that has been set on the CEL-450/490.
It is recommended that the highest setting (115200) be selected
on both the instrument and dB23 in order to give the fastest
communication.
(NOTE: When communicating with sound level meters via a GSM
modem, a Baud rate of 9600 must be used.)
No communication will be possible if the instrument and software
have different baud rates selected (See the CEL-450/490 manual
for instructions on setting the instrument baud rate).
6. Select either 1 or 2 stop bits.
When communicating with a PC it makes no difference which
option is selected.
7. Select OK to accept the settings and close the dialog.
The settings will be stored and retained for further dB23 sessions.
8. Switch the CEL-450/490 ON.
The sound level meter recognises that it is connected to a PC
and indicates this during the start up sequence, before
continuing on to a Calibration screen.
Communication with the sound level meter can now occur at any
time. However, if the instrument is carrying out a run, the
current measurement cannot be downloaded, although all other
stored runs can be accessed.
Interface status: RS 232 connected
Establishing Communication
Page 10 - CEL dB23 Users Handbook
Comentarios a estos manuales