4. SERVICE WARNINGS !
The Microdust pro and associated Environmental Enclosure is non-
intrinsically safe and must not be used within a hazardous area.
BATTERY WARNINGS !
DO NOT ENABLE the charger circuit in the Microdust pro while non-
rechargeable batteries are installed in the instrument.
DO NOT leave the main power switch ON when the system is out of use.
The Enclosure’s 12 volt lead acid battery will discharge into the
Microdust pro battery causing DAMAGE.
DO NOT use any other charger than the one supplied by CASELLA CEL.
The charger is designed for indoor operation only.
DO NOT allow the batteries to remain discharged. Permanent damage
may be done to both instrument and batteries if they are left discharged.
Re-charge as soon as possible after use.
DO NOT connect the Environmental Enclosure power supply / battery
charger directly to any Microdust Instrument. Damage to the instrument
will occur.
DISPOSAL OF BATTERIES: Batteries and
battery packs must never be disposed of by
placing in a fire or incinerator, nor must they be
punctured, crushed or otherwise mutilated or
opened up in any way.
NiCd batteries contain CADMIUM and must be
disposed of in a safe manner; in some countries
this may involve specialist licensed waste
disposal companies.
Warnings
Cd
Service