4. A Bargraph (Refer to Figure 4 [A]) or Time History (Refer to Figure 4 [B]) will be
displayed. Refer to
INSTRUMENT SETTINGS to change the Time and
Frequency Weightings, or to change the type of display.
CEL-240/242 CEL-244/246
Figure 4 CEL-200 Series Bargraph and Time History Screens
5. The screen will display the instantaneous and maximum sound pressure levels
along with the relevant Time and Frequency Weighting. Refer to
Instrument
Settings to change these settings. Additionally, CEL-244 and CEL-246
instruments will display the average level (Leq or Lavg) since the reset button
was pressed. To reset the maximum sound pressure level and average level,
press and release the
key.
6. When the required measurements have been taken, press and hold the Power
ON/OFF key (Refer to
Switch Off) for three seconds to switch the instrument
OFF.
Note: The ‘press and hold for three seconds’ feature prevents the CEL-200 Series
being accidentally switched OFF during use.
Back to Index
5. CALIBRATION
It
is
recommended
that
an
acoustic
calibration
using
a
CEL-110/2
(
or
CEL-110/1)
calibrator be performed
before
and
after
taking
measur emen ts
.
PROCEDURE
1.
Switch
ON
the
Acoustic
C
alibrator
(Refer
to
the
Calibrator
Instruction
Manual).
2. Check the
Acoustic
C
alibrator has a 114dB (or 94dB) output.
3.
Make sure the instrument is set to the 60-130dB range.
4. Fit the
Acoustic
C
alibrator firmly over the microphone. The
instrument
automatically
detects
a
1kHz
tone
from
a
94
or
114dB
Calibrator
and enters
Calibration
Mode
(
Refer to Figure 5
).
8